Check cable voltage drop and transmitter headroom on a 4-20 mA current loop — worst case at 20 mA and at the 22 mA HART diagnostic peak.
A transmitter sits 500 m from the marshalling cabinet on 1.5 mm² twisted-pair cable at 20°C. Supply is 24 V, the transmitter needs at least 12 V, and the receiving instrument burden is 250 Ω.
The form above is pre-filled with these exact numbers — press Calculate to reproduce this result, or change the inputs for your own loop.
| Field | Accepted range | Notes |
|---|---|---|
| Cable length | > 0 | m, km or ft |
| Cable cross-section | > 0 | mm² or AWG (10–28 tabulated, other gauges computed) |
| Supply voltage | > 0 | V |
| Transmitter min voltage | > 0, < supply | V |
| Loop burden | ≥ 0 | Ω |
| Cable temperature | −50 to 200°C | Defaults to 20°C if left blank |
| Cable / CSA | Headroom @ 20 mA | Status |
|---|---|---|
| 500 m / 1.5 mm² | 6.78 V | PASS at 20 mA and 22 mA |
| 5000 m / 0.5 mm² | 0.28 V | PASSES 20 mA, FAILS 22 mA diagnostic |
| 6000 m / 0.5 mm² | −1.06 V | FAILS at 20 mA |